5 | Installation
Büchi Labortechnik AG
26/102
Operation Manual Pure Chromatography Instruments
Precondition:
R
Make sure that the instrument is not
connected to the power supply.
u
Attach all solvent lines to the
instrument. Connections see
Chapter 3.2.4 "Connections on gradient
valve", page 17
u
Put the other end of the solvent line
into the solvent bottle.
u
Assign the solvent to the solvent lines.
See Chapter 5.7 "Assigning solvent to
solvent lines", page 26
u
Install the solvent level sensor. See
Chapter 5.6 "Installing the solvent level
sensor", page 26
5.6
Installing the solvent level sensor
Precondition:
R
Make sure that the instrument is not
connected to the power supply.
u
Attach all solvent level sensors to the
instrument. Connections see
Chapter 3.2.4 "Connections on gradient
valve", page 17
u
Calibrate the solvent level sensors. See
Chapter 5.8 "Calibrating the solvent
level sensor", page 27
5.7
Assigning solvent to solvent lines
Navigation path
➔
Tools
➔
Solvent Loading
Precondition:
R
The solvent bottle is connected to the instrument. See Chapter 5.5 "Establishing
solvent connection", page 25