Operation Manual Pure Chromatography Instruments
63/102
Possibilities to edit the gradient during a separation:
Change the gradient. See Chapter "Editing the gradient in graphic mode", page 47
Hold the gradient. See Chapter 6.1.6 "Gradient panel", page 39
Auto gradient hold. See Chapter 6.1.6 "Gradient panel", page 39
6.5.5
Ending a Flash separation
NOTE
The separation extends automatically by 5 minutes if the baseline at the end of the
separation is not below the set threshold.
Precondition:
R
The display shows the dialog box
Separation End
.
u
According to the requirements extend the separation time by 5 minutes.
Precondition:
R
The display shows the dialog box
Run completed
.
u
Follow the instructions on the display.
ð
Purging the instrument with air. See Chapter 7.8 "Removing solvent from an
used cartridge", page 78
ð
Removing the cartridge. See Chapter 6.4.3 "Installing and removing a cartridge",
page 57
6.5.6
Shutting down the instrument
Navigation path
➔
File
Precondition:
R
The separation process has ended.
u
Purge the instrument with purging solvent. See Chapter 7.6 "Cleaning the
instrument", page 77
u
Navigate to the
[File]
menu via the navigation path.
u
Tap the action
[Shut down]
.
u
Confirm the secure question with
[Yes]
.
ð
The instrument is shutting down.
6.6
Performing a Flash separation manually
NOTE
Pure C-810 / C-815 / C-850 only
6.6.1
Preparing the instrument
Time required:
approx. 30 sec.
Precondition:
R
All commissioning operations have been completed. See Chapter 5 "Installation",
page 24
u
Switch the
On/Off
master switch to On.