Installation | 5
Operation Manual Pure Chromatography Instruments
25/102
5.4
Establishing electrical connections
NOTE
Observe the legal requirements when connecting the instrument to the power
supply.
u
Use additional electrical safety features (e.g., residual-current circuit breakers) to
comply with local laws and regulations.
The power supply must fulfil the following conditions:
1. Provide the mains voltage and frequency specified on the type plate of the
instrument.
2. Be designed for the load imposed by the instruments connected.
3. Be equipped with suitable fuses and electrical safety features.
4. Be equipped with proper earthing.
NOTICE
Risk of property damage and diminished performance due to use of unsuitable
power cables.
The power supply cables supplied with the product by BUCHI precisely match the
requirements of the device. If other power cables that do not meet those
requirements are used, the device may be damaged and/or its performance
diminished.
u
Use only the power supply cables supplied with the product or ordered
separately from BUCHI.
u
If using any other power supply cables, make sure that they match the
specifications on the type plate.
u
Make sure that all connected devices are earthed.
u
Plug the power cable into the connection on the instrument. See Chapter 3.2
"Configuration", page 14
u
Plug the mains plug into the mains outlet socket.
5.5
Establishing solvent connection
NOTICE
Solvent bottles on top of the instrument.
Solvent bottles on top of the instrument can cause property damages.
u
Locate the solvent bottles next to the instrument.
u
Use the optional solvent bottle platform.