7 | Cleaning and servicing
Büchi Labortechnik AG
74/102
Operation Manual Pure Chromatography Instruments
Component
Action
Frequency
Nebulizer
u
Clean the nebulizer. See Chapter 7.4
"Cleaning the nebulizer", page 75
Monthly
Air filter
u
Replace the air filter.
Yearly
7.2
Carrying out a data backup
Navigation path
➔
File
➔
Exit
u
Navigate to the
Exit
dialog according to the navigation path.
u
Confirm the secure question.
ð
The Pure software is shutting down.
ð
The display shows a Windows® system.
u
Connect a USB storage device to the instrument.
u
Open the Windows® Explorer.
u
Navigate to the data you wish to backup. See Chapter 10.2 "Folder locations",
page 96
u
Copy the needed data to the USB storage devices.
7.3
Calibrating the display
Navigation path
➔
Tools
➔
Calibration and Defaults
u
Navigate to the
Calibration and Defaults
dialog according to the navigation
path.
u
Tap the button
[Calibrate]
.
u
Follow the instructions on the display.