7 | Cleaning and servicing
Büchi Labortechnik AG
78/102
Operation Manual Pure Chromatography Instruments
u
Navigate to the
Column Flushing
dialog according to the navigation path.
u
Enter the required data according to your needs.
7.7
Cleaning the solid loader
Navigation path
➔
Tools
➔
Manual Control
➔
Solid Loader Flushing...
u
Install a bypass. See Chapter 6.4.4 "Installing and removing a column", page 58
u
Navigate to the
Solid Loader Flushing
dialog according to the navigation
path.
7.8
Removing solvent from an used cartridge
Navigation path
➔
Tools
➔
Manual Control
➔
Air Purging
Precondition:
R
The cartridge is installed.
u
Navigate to the
Air Purging
dialog according to the navigation path.
u
Enter the purging time according to your needs.
u
Tap the button
[Column purge]
.
ð
The instrument and the cartridges are cleaned.