Büchi Labortechnik AG
Installation | 5
Operation Manual Pure Chromatography Instruments
27/102
R
The solvent you wish to use is part of the solvent library. See Chapter 6.3 "Editing
a solvent", page 53
u
Navigate to the
Solvent Loading
dialog according to the navigation path.
ð
The display shows the dialog box
Solvent Loading
.
u
Tap the drop-down list next to Line 1.
ð
The display shows a drop-down list with selectable solvents.
u
Select the solvent which is connected to solvent line 1.
ð
The solvent for Line 1 is assigned.
ð
The drop-down list closes.
u
Repeat the solvent selection for each line.
u
Activate the checkbox next to the line you wish to prime.
u
Tap the button
[Auto Prime]
.
u
Wait till priming finished.
u
Tap the button
[Close]
.
ð
All lines are assigned with solvents.
ð
The dialog box closes.
5.8
Calibrating the solvent level sensor
Navigation path
➔
Tools
➔
Calibration and Defaults
Precondition:
R
The solvent level sensor is not covered by solvent.
u
Navigate to the
Calibration
dialog according to the navigation path.
ð
The display shows the dialog box
Calibration
.
u
Tap
[Zero]
for the related solvent line.
u
Tap the button
[Close]
.
ð
The dialog box closes.