PAT I E N T S A F E T Y
PATIENT SAFETY
Read Instructions
for Use prior to use.
The infusion device
should only be
used by specially
trained staff.
Intended use
The Infusomat® Space P Volumetric Infusion Pump System includes an external
transportable electronic volumetric infusion pump, dedicated administration
sets, and pump accessories. The system is intended for use on adults, pediatrics,
and neonates for the intermittent or continuous delivery of parenteral and
enteral fluids through clinically accepted routes of administration. These routes
include, but are not limited to intravenous, irrigation/ablation, and enteral. The
system is used for the delivery of medications indicated for infusion therapy
including but not limited to colloids and cristalloids, blood and blood compo-
nents, Total Parenteral Nutrition (TPN), lipids, and enteral fluids. The Infusomat®
Space P Volumetric Infusion Pump System is intended to be used by trained
healthcare professionals in healthcare facilities, home care, outpatient, and
medical transport environments.
Qualified medical staff should decide how the device should be used based on
its features and specifications. For more details, please read the Instructions for
Use.
Operation
•
The initial training of the Infusomat® Space P is to be performed by B. Braun
sales personnel or other authorized persons. After each software update, the
user is required to inform himself about the changes to the device and
accessories in the instructions for use.
4
w
Caution:
Ensure the unit is properly positioned and secured. Do not position
pump unit above patient or in a position where a patient could
come to harm, should the pump fall.
•
Prior to administration, visibly inspect the pump for damage, missing parts or
contamination and check audible and visible alarms during selftest.
•
Not be used adjacent and stacked with other equipment except B. Braun
Space devices.
•
Only connect to patient once the line has been correctly inserted and
completley primed. Interrupt connection during line change to prevent
incorrect dose delivery.
•
During priming and bolusing the pressure limits are set to the maximum level.
•
Select infusion line/catheter suitable for use with the intended medical
application.
4
w
Caution:
Position the infusion line free of kinks.
•
Recommended change of disposable every 24 h (or as per national hygiene
regulations).
w
6