51
"Flow" Drop chamber is completely filled or leak in the
system.
Examine the line for damage and check the drop
chamber.
“Data were reset” Therapy and pump settings could not be restored.
Enter therapy again.
"Therapy data were reset" Therapy data could not be restored.
Enter therapy again.
“Data Lock" An attempt was made to stop or switch the pump off
without entering the code.
Enter the correct code in order to continue the
therapy respectively turning the pump off.
The red LED extinguishes with the acknowledgement of the alarm.
Caution:
If a wrench (
) is displayed and/or a yellow, red and blue LED blink
then the pump is in the service mode and is not permitted to be used on a patient. The
pump must then be checked by a service technician.
Caution:
If the
is constantly displayed in the headline, the audible alarm for the
pre- and operating alarm is silenced for a predefined time via service tool. Only the
visual alarm is still displayed at the pump. After the predefined time is elapsed the
pump provides the audible alarm.
6.3 Reminder Alarms
Reminder alarms only occur in two cases:
1. A line is inserted, the pump does not deliver, no value is edited and the device is not
operated for two minutes.
An acoustic tone sounds, the yellow LED is constantly on and a staff call is
activated.
a) The display states “Reminder alarm!”
b) The display states “Config. not finished!”
Confirm alarm with
K
and continue to set therapy/Start Up configuration.
2. A value edition was started but not finished and confirmed. This is also possible
with a missing disposable.
An acoustic tone sounds, the display states “Value not accepted“, the yellow LED
is constantly on and a staff call is activated.
Confirm alarm with
K
and continue to set therapy.
ALARMS
Chapter 6