35
The Main menu informs about the current interval. The configured parameters
can be checked by Program Parameter Menu in Main.
3.9 Intermittent Mode
The Intermittent Mode consists of 2 phases. This phases will be repeated.
•
Bolus phase: the configured bolus is active
•
Rate phase: time during the therapy in which the entered rate is active
Example:
Intermittent Mode should only be performed by an experienced user being familiar
with the principles of the Intermittent Mode and properly trained in using the present
device.
Note:
The active Multi Dose Mode function always displays this icon in the Display
(
).
Note:
Regular Bolus function is disabled for Intermittent Mode.
In Intemittent Bolus the bolus service settings are active. The pressure level is auto-
matically set to max value.
Set Profile Parameters: The therapy can be started directly via the drug library or via
the Main Menu/Special functions.
Starting Intermittent Mode via Drug Library:
Note:
Intermittent Mode settings have been configured in the Drug List
Manager before and have been uploaded into the pump.
•
Switch on pump with
o
and wait until self-check is finished.
•
Insert disposable and use the drug library according to Instructions for Use.
SPECIAL FUNCTIONS
Chapter 3