ALARMS
Chapter 6
ALARMS
The Infusomat® Space P is equipped with an audible and optical alarm signal.
6.1 Device Alarms
When a device alarm occurs the infusion is immediately stopped. Press
o
to switch
off the device. Then switch the device on again. In case of a repeated device alarm you
must close the rollerclamp, disconnect from the patient, open the front door of the
pump and take out the disposable. The device needs to be handed to the service
department.
6.2 Pre-Alarms and Operating Alarms
Pre-alarms:
Pre-alarms occur a few minutes (dependable on service settings) prior to operating
alarms. During pre-alarms an audible tone sounds, the yellow LED is constantly on and
a staff call is activated (optional). The display message varies depending on the alarm
reason. The signal tone and the staff call are turned off with
C
. Display and LED stay
in pre-alarm until the operating alarm goes off. Pre-alarms don't lead to an
interruption of the infusion.
Alarm-
type
Device
Alarm
yes
flashes
off
device alarm
and alarm
code
(see service
program)
yes
Press
O
and follow
the instruction on the
display.
Opera-
tingA-
larm
yes
flashes
off
see alarm de-
scription
yes
Pre-
Alarm
yes
off
constant
on
see alarm de-
scription
(de-)activate
via service
program
Press
C
to mute
alarm and turn off
staff call. Visible
alarm remains until
end.
Remin-
der
Alarm
yes
off
constant
on
see alarm de-
scription
yes
Press
k
to mute
alarm, turn off staff
call and delete the
alarm text.
Alarm
Hint
no
off
off
see alarm de-
scription
no
Hint disappears
without
confirmation.
Audible
signal
Red LED
Optical signal
Yellow LED Text
Staff call
User confirmation
Press
k
to
acknowledge the
audible alarm, alarm
text and staff call.
Press
C
to mute the
alarm for 2 minutes.