Chapter 3
SPECIAL FUNCTIONS
The settings for the steps of the infusion are shown on the display. These
settings, configured in the Drug List Editor, need to be confirmed with
r
.
•
To change the values, press
l
to change and
k
to confirm.
•
Adjust VTBI with
q
.
The pump can be started now by pressing
s
f
.
Starting Program Mode via Special Function Menu:
•
Switch on pump with
o
and wait until self-check is finished.
•
Insert disposable.
•
Go to Special Functions Menu and select Program Mode.
•
Press
l
to enter parameters and
k
to confirm.
•
Adjust VTBI with
q
.
After entering all desired parameters the pump can be started by pressing
s
f
.
In the upper part of the display the icon for “Program Mode” appears.
The screen shows the following:
Only the VTBI may be changed during an infusion that is running.
•
Press
C
to check upcoming Program Mode intervals in Main Menu.
It is possible to cancel one step of the running therapy. All following steps in
the programmed sequence persist.
•
Go to Main Menu by pressing
C
.
•
Use
t
to navigate through the Main Menu and select Current with
l
.
•
For checking upcoming intervals press
C
.
•
Select "Program Parameters" with
l
.
•
Go through all interval steps with
r
.
Note:
The delivery of drugs can be stopped and started again in the Program
Mode at any time by pressing
s
f
.
Number of cycles is defined by VTBI. Take care to set the VTBI in the correct
relation to the volume of one Cycle. VTBI may needs to be adjusted after
changing the intervals.
The pump infuses the
predefined rate in the
predefined time for the current
step.