INFUSOMAT® SPACE P OVERVIEW
4
Transport
A maximum of three pumps (Infusomat®
Space P or Perfusor® Space) plus one
SpaceControl may be stacked together (in
ambulance cars and helicopters only one
pump). Avoid external mechanical
influence.
Locking Devices Together
Line up the bar of the lower pump with the
bar of the pump above and slide the lower
pump backwards until the lock clicks and
the green buttons are above each other.
To disconnect, push green locking buttons
of top pump device and slide bottom pump
forward.
Caution:
Avoid external mechanical action.
Pole Fixation
Push the opening of PoleClamp against the
vertical pole and lock the screw tightly.
Unscrew to release.
For vertical position push lever down and
rotate either way until lever clicks into
notch. Push lever for
rotation.
Caution:
A maximum of three B. Braun
Space pumps can be stacked together
only in horizontal pump position
when used with the PoleClamp SP.
Pump based free flow
prevention safety clamp
Downstream
pressure sensor
Upstream
pressure sensor
Line guide
cover
Air
sensor
Fixaton of PoleClamp (Universal Clamp)
Line up bar of pump with bar of PoleClamp
and slide PoleClamp forward until locking
mechanism clicks.
To remove, press release button on frame,
push handle down and pull PoleClamp
backwards.
Pole clamp handle
Pole clamp
release button
Pump slots
Yellow
caution light
Opening
lever
Green locking
button