21
SPECIAL FUNCTIONS
ADVANCED OPERATIONS
2.1 Status Request of Pump when Infusion is Running
Press
c
to switch between run display and Main Menu while the device is
infusing. Navigate through the menu using
t
to check parameters. In order to
check the menu parameters in the Status-/Options Menu, select "Status" res-
pectively "Options" in the Main Menu, open menu with
l
and scroll through
menu with
t
.
2.2 Rate, VTBI and Time Change Without Infusion
Interruption and Reset Of Status Menu Data
•
Press
c
when the pump is in the run display in order to switch to the Main Menu.
Select rate/VTBI/time with
t
and press
l
in order to open the parameter.
•
Enter new value with
q
and confirm with
k
.
Reset Status Menu Data:
The parameters intermediate volume and -time can be reset when the pump is
infusing or when the pump has stopped.
•
Select “Status” in Main Menu with
t
and press
l
.
•
Highlight intermediate volume (in ml) or intermediate time (in h:min) with
t
and
open parameter with
l
.
•
Reset values by pressing
u
.
Both parameter total volume and -time, are displayed in the pump as "Total" with the
according unit and can be reset by starting a new therapy. A second way to reset the
parameters when the pump is in the Main Menu: Press
c
, answer the question
whether the last therapy is to be used with
u
and reset the values with
u
.
The type of the inserted line is displayed in menu item „Line“ and cannot be changed
once it has been confirmed at the beginning of the infusion. The drug info states the
drug name, the name of the drug list and its date of origin. If the change from the
secondary to the primary infusion will be performed manually or automatically will be
displayed in line “PGY change”. The current battery capacity in hours and minutes is
displayed in the menu item “Battery Cap.” and the current software version in menu
item "Version". In-line pressure can also be read in the Status menu in mmHg or Bar
depending on the service settings.
Chapter 2