Note:
The feature Dose Over Time always requires the usage of dosing units (i.e.,
mg or mg/kg patient weight).
Before using Dose Over Time contact your local B. Braun representative!
Starting Dose Over Time via Drug Library:
Note:
Dose Over Time settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
•
Switch on pump with
o
and wait until self-check is finished.
•
Insert disposable and use the drug library according to the Instructions for
Use.
•
Select a drug by using
t
and press
l
.
The pump now offers the possible therapy profiles. Select “Dose over Time” with
t
and
press
l
.
The editor for Total Dose is shown if a drug with therapy Dose over Time is sel-
ected from drug library and no default value for Total Dose was entered in libr-
ary. The editor is also shown if the Total Dose is edited in the Main menu.
The editor for Time is shown if a drug with therapy Dose over Time is selected
from drug library and no default value for Time was entered in library. The edi-
tor is also shown if the Time is edited in Main Menu.
The VTBI is calculated automatically and the following screen is displayed:
Note:
It is always possible to press the key
C
in the Run Menu and edit or
check values in the Main Menu while the pump is delivering.
Enter the total dose, if necessary,
and confirm with
k
.
Enter the time, if necessary,
and confirm with
k
.
Check calculated rate by using
t
for plausibility
Start Dose Over Time by
pressing
s
f
.
Run Menu: The time is used to
control the therapy. For this
reason the remaining time is
shown big digits in menu Run.
The parameter in the lower left
corner can be scrolled. Set to
Rate when leaving the pump.
Chapter 3
SPECIAL FUNCTIONS