33
3.8 Program Mode
Program Mode is for infusion requiring a non-standard delivery pattern. The user de-
fines a series of intervals (max. 12 intervals) by certain parameters (rate, time, volume)
for each cycle.
The pump automatically gives each programmed period, one after the other
Example:
Program Mode should only be performed by an experienced user being familiar with
the principles of the Program Mode function and properly trained in using the present
device.
Note:
The active Program Mode function always displays this icon in the Display
(
).
Note:
Bolus function is disabled for Program Mode.
Set Profile Parameters: The therapy can be started directly via the drug library or via
the Main Menu/Special functions.
Starting Program Mode via Drug Library:
Note:
Program Mode settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
•
Switch on pump with
o
and wait until self-check is finished.
•
Insert disposable and use the drug library according to Instructions for Use.
•
Select the desired drug with
t
and press
l
.
•
Select Program Mode with
l
.
In the following screen the user has to confirm the number of steps for the
therapy with
k
.
SPECIAL FUNCTIONS
Chapter 3