Braun Infusomat Space P Instructions For Use Manual Download Page 32

32

Chapter  3

SPECIAL FUNCTIONS

Taper phase

Note: 

After starting infusion it is only possible to change rates, time and VTBI in the

continuous phase. 
By editing (increasing/decreasing) the plateau rate, the therapy is recalculated. With
the increase/decrease of the plateau rate the volumes in the ramp phase, the conti-
nuous phase and the taper phase are increased/decreased. The continuous phase is
shortened/prolonged to infuse the VTBI still completely with the end of the taper 
phase. 

By editing the Ramp/Taper-Time, the therapy is recalculated. The Continuous Phase is
extended/shortened to infuse the VTBI still completely until the end of the Taper 
phase.

By increasing/decreasing the VTBI, the continuous phase is prolonged/shortened to in-
fuse the new entered VTBI completely with the end of the taper phase.

Note:

The delivery of drugs can be stopped and started again in Ramp and Taper Mode

at any time bypressig 

s

f

. Ramp and Taper is stopped immediately without Taper phase

and started without a new Ramp phase. This will not have any effect on the settings of
the therapy.

Immediate Taper Down

By chosing the Immediate Taper Down Function the therapy can be ended with a taper
phase before the originally defined VTBI is completely infused.

   Press 

C

during continuous phase.

   Use 

t

to select Special Functions and press 

l

.

   Select Immediate Taper Down Function and confirm with 

l

.

   Edit taper time by using 

t

and press 

k

to confirm. 

      The pump automatically changes to Taper phase and linearly decreases the 
      rate.

The pump linearly decreases the
rate in the predefined time
until it reaches the KVO rate

Summary of Contents for Infusomat Space P

Page 1: ...s Instructions for Use Valid for software 687N GB It is recommended that all pumps at your care unit are equipped with the same software version or in parallel use with Software M 38911866N_Infusomat...

Page 2: ...algesia PCA 26 3 5 Barcoding 27 3 6 Piggyback Function 28 3 7 Ramp and Taper Mode 30 3 8 Programm Mode 33 3 9 Intermittent Mode 35 3 10 Dose Over Time 37 Chapter 4 Autoprogramming 39 Chapter 5 Options...

Page 3: ...ment Before changing the battery always disconnect the pump from the patient and switch off the device To remove the battery cover push the button below the battery compartment with a pointed pen and...

Page 4: ...eClamp against the vertical pole and lock the screw tightly Unscrew to release For vertical position push lever down and rotate either way until lever clicks into notch Push lever for rotation Caution...

Page 5: ...5 SYMBOLS ON PRODUCT SYMBOLS ON PRODUCT...

Page 6: ...ow the device should be used based on its features and specifications For more details please read the Instructions for Use Operation The initial training of the Infusomat Space P is to be performed b...

Page 7: ...ing must be carried out if life saving medication is performed The air detector cannot detect air diffusing in the following components three way stopcocks infusion adapters and further lines placed b...

Page 8: ...01 1 2 2007 and IEC 60601 2 24 2012 are maintained If the equipment is operated in the vicinity of other equipment which may cause high levels of interference e g HF surgical equipment nuclear spin to...

Page 9: ...Y Safety Instructions for using Pole Clamp 1 Line pump up with the Pole Clamp guide rails 2 Slide pump fully into place onto the guide rails 3 An audible Click should heard 4 Test the pump is secure W...

Page 10: ...NAVIGATION MENU STRUCTURE NAVIGATION All display screen shots are examples and may be different when related to an individual patient and individualized therapy o On Off button x Door open button s f...

Page 11: ...ear end an audible tone andthe yellow LED is con stantly on To confirm a pre alarm press k This hint pops up if a user tries to edit or change a parameter by pressing l when that parameter is unable t...

Page 12: ...Dose Drug Library Change over from continuous mode to PCA Intermediate volume Intermediate amount Intermediate time Total volume Special Functions Options Status Options Menu Occlusion Pressure Total...

Page 13: ...nd u to open the pump door in order to continue with inserting the line 4 w Caution Close the roller clamp before inserting the IV line and do not connect to patient until properly loaded and primed C...

Page 14: ...pull the door shut Do not open roller clamp until the pump directs you to do so when self test is completed Then select the inserted line with t and confirm it with l Open the roller clamp Caution Bef...

Page 15: ...ed to start the infusion Time will be calculated when rate is entered When time is entered rate doserate will be calculated Running arrows on the display and the green LED indicate that the pump is in...

Page 16: ...ymbol apparent in the Main Menu shows the established para meter for the application VTBI or time When the rate is changed the so called target parameter is principally not adjusted to the new rate bu...

Page 17: ...fter finishing the bolus volume 3 Bolus with rate calculation Press n b Then press l and set bolus dose by using q Press k to confirm bolus dose Set time with q in which a bolus is to be delivered Cal...

Page 18: ...connection Press x and open the pump door with u Press down the green opening lever completely until it locks in place remove the line and insert a new line Note In the unlikely event the pump door ca...

Page 19: ...and follow the instructions as described 1 5 End of Infusion Press s f to stop the infusion The green LED goes out Close the roller clamp and interrupt the patient connection Press x Answer the questi...

Page 20: ...is supposed to switch to standby by pressing u The pump is now in Standby While the pump is in the standby mode it s display shows the drug and the remaining time for this mode Change of remaining ti...

Page 21: ...ighlight intermediate volume in ml or intermediate time in h min with t and open parameter with l Reset values by pressing u Both parameter total volume and time are displayed in the pump as Total wit...

Page 22: ...before starting the infusion also if body surface area related dose rate is set by Barcode The dose rate calculation enables a calculation of the rate in ml h based on the entered dose parameters Set...

Page 23: ...A deactivation of the dose rate calculation is only possible when the pump is stopped Press c from Main Menu and then press d Caution A change of the patient weight or height will alter the flow rate...

Page 24: ...a drug choose therapy type with t and confirm with l Confirm the displayed drug information with l Decide if the safety limits for the drug are to be applied u or if only the drug name should be used...

Page 25: ...s within the range of the minimum soft limit Violation of the upper soft limit Violation of the lower soft limit No soft limit is defined Only a drug name is available It is possible to select a drug...

Page 26: ...lly in case non authorised access to the settings can be anticipated The code is entered with q and confirmed with k In this state the patient is allowed to demand boli Depending on the status of the...

Page 27: ...PCA Chart is displayed The bar represents a time axis with the points above the axis representing the number of boli administered and the points below the axis representing the number of boli refused...

Page 28: ...can be done manually or automatically Correspondingly if an automatic change is to be made automatically or manually is to be answered with u or d The rate and VTBI of the secondary infusion can be l...

Page 29: ...cament Go to Set new Piggyback in the Special Functions Menu and confirm with l Note The Piggyback infusion can be deleted by starting a new Piggyback and by pressing cancel and answering the question...

Page 30: ...teau rate is used as a continuous infusion Taper phase the pump rate is decreased linearly after the continuous phase until the KVO rate is reached or pump is stopped in a predefined time Down Time Ex...

Page 31: ...rted now by pressing s f Starting Ramp and Taper via Special Function Menu Switch on pump with o and wait until self check is finished Insert disposable Go to Special Functions Menu and select Ramp an...

Page 32: ...nuous phase is prolonged shortened to in fuse the new entered VTBI completely with the end of the taper phase Note The delivery of drugs can be stopped and started again in Ramp and Taper Mode at any...

Page 33: ...n always displays this icon in the Display Note Bolus function is disabled for Program Mode Set Profile Parameters The therapy can be started directly via the drug library or via the Main Menu Special...

Page 34: ...or Program Mode appears The screen shows the following Only the VTBI may be changed during an infusion that is running Press C to check upcoming Program Mode intervals in Main Menu It is possible to c...

Page 35: ...evice Note The active Multi Dose Mode function always displays this icon in the Display Note Regular Bolus function is disabled for Intermittent Mode In Intemittent Bolus the bolus service settings ar...

Page 36: ...ter parameters and k to confirm After entering all desired parameters the pump can be started by pressing s f In the upper part of the display the icon for Intermittent Mode appears In bolus phase the...

Page 37: ...lised with a characteristical symbol in the Display If besides DOT the piggyback therapy is active a combined symbol for both therapies will be displayed Note Dose Over Time should only be performed b...

Page 38: ...l Dose was entered in libr ary The editor is also shown if the Total Dose is edited in the Main menu The editor for Time is shown if a drug with therapy Dose over Time is selected from drug library an...

Page 39: ...seconds Ensure pump is in the Main Menu passive mode or Standby New Order message will appear with drug name and mode Press k to accept or c key to cancel order and respond to prompt Select Care Unit...

Page 40: ...first order sent send as Continuous is always considered as the PRIMary infusion subsequent orders will be considered PIGGYback Note Order sent as Dose over Time is always considered the PRIMary infus...

Page 41: ...o clear current infusion PIGGYback Orders Orders received after PRIMary has been set will be for PIGGYback infusions follow prompts on screen to stop the PRIMary to accept the PIGGYback order Confirm...

Page 42: ...scroll to any value and open editor with l to change value Alternately order may be cancelled and request made for revised order to be sent Note If pump is placed in standby while order is pending ne...

Page 43: ...influences such as the tube length tube diameter viscosi ty of the liquid and the filter used in the system set up adjustments may need to be made to the pressure level Enter pressure in Options Menu...

Page 44: ...se between nine pressure levels 1 lowest level 9 highest level by pressing l or r and confirm entry with k 5 3 Data Lock The data lock function protects the device against unauthorized access It is re...

Page 45: ...by the following table Activation of the function Open data lock in Options Menu with l Select between level 1 2 or 3 if activated with l and r and confirm with k Enter code with q and press k in orde...

Page 46: ...ion of the KVO delivery is selected in the service program Open the KVO mode in the Options Menu with l Answer the Yes No question with u to enable the KVO mode 5 6 Contrast Display Light Keypad Light...

Page 47: ...Answer Yes No question by pressing u to activate the macro mode For quick activation of macro mode Press and hold r while the pump is infusing until the font size changes 5 10 Language This function e...

Page 48: ...son The signal tone and the staff call are turned off with C Display and LED stay in pre alarm until the operating alarm goes off Pre alarms don t lead to an interruption of the infusion Alarm type De...

Page 49: ...sage will remain on the screen until it is confirmed by pressing k After 2 minutes if the alarm has not been confirmed by pressing k the audible tone will sound again The alarm tone the alarm message...

Page 50: ...py Check upstream The upstream sensor triggers an alarm Check if roller clamp is closed or infusion line is kinked If a drop sensor is connected the upstream alarm is deactivated Air bubble Accumulate...

Page 51: ...displayed in the headline the audible alarm for the pre and operating alarm is silenced for a predefined time via service tool Only the visual alarm is still displayed at the pump After the predefined...

Page 52: ...oper entries are made the display states corresponding hints e g Bol rate out of range Download failed The parameter can not be modified These hints disappear after a few seconds and don t need to be...

Page 53: ...ges in the delivery rate may affect the battery operating time Disposable change procedures require a high power consumption A sudden break down of the battery operating time can be possible with an a...

Page 54: ...enance it can happen that the battery pre alarm time can no longer be maintained In this case it is necessary to perform a battery maintenance To initiate the discharge process the message Battery mai...

Page 55: ...nge IV set specification significant to system accuracy without prior notification System accuracy is 5 typical by volume as measured using the trumpet curve test method defined in IEC 60601 2 24 at r...

Page 56: ...ime of operation 100 continuous operation Operating conditions Relative humidity 30 90 without condensation Temperature 18 35 C 64 F 95 F Atmospheric pressure 500 1060 mbar Storage conditions Relative...

Page 57: ...ng multiple infusion lines with patient port different flow rates may affect the rate for all infusions past the point of connection Rate increments 0 1 99 99 ml h in increments of 0 01 ml h 100 0 999...

Page 58: ...y lead to higher emission or reduced immunity Use only compatible combinations of equipment accessories working parts and disposables with luer lock connectors Essential Performance for Infusion pumps...

Page 59: ...sed RF energy is transmitted by the Space System Refer to technical data of Battery Pack SP with Wifi IFU and or SpaceStation and SpaceCom for details RF emissions CISPR 11 Harmonic emissions IEC 6100...

Page 60: ...s power quality should be that of a typical commercial or hospital environment Surge according IEC 61000 4 5 Gegentaktspannung 1KV Gleichtaktspannung 2KV 1KV 2KV Mains power quality should be that of...

Page 61: ...in watts W according to the transmit ter manufacturer and d is the rec ommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surv...

Page 62: ...the recommended separation distance for transmitters in the frequency range 0 15 MHz to 2 5 GHz to decrease the likelihood that mobile portable communications equipment could cause interference if it...

Page 63: ...m the date of delivery for every Infusomat Space P 12 months for every Battery Pack SP This covers repair or replacement of parts damaged as a result of design manufacturing errors or material defects...

Page 64: ...The housing of Infusomat Space the slide guide and the inner tube guiding areas may be cleaned with EPA regi stered hospital disinfectants containing 1 propanol or didecyl dimenthyl ammonium chloride...

Page 65: ...itive electrical components and or electrical shock If fluid exposure occurs the device should be swapped out with another device in a manner that presents minimal interruption to patient care The dev...

Page 66: ...sider the applicable hygiene and disposal regulations Inspection on Delivery Despite careful packaging the risk of damage during transport cannot be entirely prevented Upon delivery please check that...

Page 67: ...ce P and Patient Safety Power Supply SP 8713110D 8713123D The Power Supply SP is adequate to supply power for a single pump and one SpaceControl 1 Connect plug of Power Supply SP with socket P2 on bac...

Page 68: ...o the aperture on the PoleClamp make sure that it audibly locks in 3 To remove the short stand Press the white button at the lower end of the PoleClamp and remove the short stand Note Use only one inf...

Page 69: ...Note A maximum of three plugs can be stacked upon each other in socket P2 Connection Lead SP 12 V 8713231 Install the Connection Lead SP 12 V in the following way 1 Connect plug to socket P2 on back o...

Page 70: ...ll Note A maximum of three plugs can be stacked upon each other in socket P2 turned off Alarm Operation Alarm Operation Alarm Operation turned on 1 sec 1 sec 1 sec turned off static without Off Alarm...

Page 71: ...tape for fixation of the demand button at the patient s arm Line fixation connection between hook and loop tape and demand button Metal clip alternatively for fixation at the bed sheet Cable strap for...

Page 72: ...133 Drop Sensor SP 8713175 Short Stand SP 8713135 Battery Pack SP NiMH 8713180 Battery Pack SP NiMH incl Pin 8713180A Battery Pack SP LiIon incl Pin and WiFi 8713182A Interface Lead CAN SP 8713230 Int...

Page 73: ...cock 220 cm 4063006 Intrafix Safeset w y needlefree valve Caresite 210 cm 4063004C Intrafix Safeset w y needlefree valve backcheck valve 210 cm 4063004KW Full list of Infusomat Space P Lines on reques...

Page 74: ......

Page 75: ......

Page 76: ...free 0318 B Braun Melsungen AG Sparte Hospital Care 34209 Melsungen Germany Tel 49 0 56 61 71 0 Fax 49 0 56 61 71 20 44 www bbraun com Manufactured by B Braun Melsungen AG 34209 Melsungen Germany Tel...

Reviews: