18
OPERATION
1.4 Infusion Line Change and New Therapy Start
Note:
Always interrupt the patient connection before changing a line to avoid dosing
errors. Never let the pump run unattended when changing the line. Check and clean
the safety clamp regularly.
•
Press
s
f
to stop the delivery. The green LED goes out. Close the roller clamp
and interrupt the patient connection.
•
Press
x
and open the pump door with
u
. Press down the green opening lever
completely until it locks in place, remove the line and insert a new line.
Note:
In the unlikely event the pump door cannot be opened remove the allen key
from the inside of the battery compartment cover. Use this key to remove the
emergency aperture cover of the pump. Place the crank in the aperture and turn it
clockwise until the pump door opens.
Push cover opening with pen.
Remove crank from inside of
battery cover.
Turn crank to remove emergency
aperture cover.
Chapter 1