TECHNICAL DATA
Sensitivity upstream sensor 9 levels from -120 mbar to -200 mbar
(pressure reduction)
Occlusion alarm pressures 9 levels up to 1.2 bar
Max. bolus after bolus reduction
≤
0.2 ml
Alarm volume 9 levels from 1 (59dBA) to 9 (74dBA)
Mechanical occlusion pressure limit
under fault conditions Occlusion alarm pressure max. 3 bar
(300 kPa).
Max. bolus volume 2 ml.
History protocol < 3000 last history entries.
100 events for system diagnose.
Refer to separate documents of the
History Viewer for closer information.
•
Use only pressure proof and compatible disposable items (min. 2 bar/1500 mm Hg)
to avoid influencing performance data - which would result in impairing patient
safety.
•
Only use combined with approved devices/accessories by the manufacturer,
otherwise this may lead to higher emission or reduced immunity.
•
Use only compatible combinations of equipment, accessories, working parts
and disposables with luer lock connectors.
Essential Performance for Infusion pumps:
•
Infusion of liquids without variation of infusion rate
•
Pressure limitation as protection from the bursting of the infusion line
•
Protection from air-infusion
•
Protection against unintended bolus volumes and occlusion (added by IEC
60601-2-24)
•
Alarm signal of high priority (added by IEC 60601-2-24)
Note:
The technical data stated in this Instructions for Use manual were
determined with the Infusomat® Space P lines as of Intrafix® Primeline Classic
(406 2957). These technical data can change when using set configurations.
00:24
00:15
00:07
[100 ml/h]
02:06
01:14
00:33
[25 ml/h]
46:50
25:53
typ. 1.2
typ. 0.7
09:07
typ. 0.3
[1 ml/h]
Occlusion pressure
Time to occlusion alarm [min] at rate
[bar]
Level 9
Level 5
Level 1
Note:
At a rate
of 0,01ml/h, the
time of occlu-
sion alarm is
> 3 hours.