Loading a drug (including the according parameters) from the Main Menu:
•
Go to Special Functions Menu and press
l
.
•
Open the drug library by pressing
l
.
•
Navigate through the list with
t
and select the care unit with
l
. If you
have already set the care unit once on your pump this step will be skipped for
the next time.
•
Change the care unit by navigating through the list until "Change Care
Area" will be displayed. Press
k
to change the care unit.
•
Navigate through the list with
t
and select the patient profile with
l
.
If no profile is set, this step will be skipped.
•
Navigate through the list with
t
and select in alphabetical order (all drugs)
or within a category with
l
.
•
If different therapies are related to a drug, choose therapy type with
t
and
confirm with
l
.
•
Confirm the displayed drug information with
l
.
•
Decide if the safety limits for the drug are to be applied
u
or if only the
drug name should be used
d
.
•
Check if the drug short name in the Run Menu is the same as the selected
drug. Check the parameter in the Main Menu with
t
and start infusion with
s
f
.
Note:
If a drug name has been assigned without safety limits, the following hint
is provided in the RUN screen:
Note:
Care unit and Patient Profile can not be changed within a therapy (incl.
Piggyback mode).
Initial Bolus:
Initial Bolus has to be configured in the Drug List Manager.
•
Use the drug library according to the instructions for use.
•
Select the desired drug with
t
and press
l
.
Before the initial bolus begins, the bolus menu is displayed to allow editing
the bolus with
q
.
•
Check the parameter and start infusion with
s
f
.
Chapter 3