27
Chapter 3
SPECIAL FUNCTIONS
An acoustic confirmation of demanded boli can be activated and modulated by
t
in Data Lock 3.
Is a demand button connected, the therapy symbol looks like this:
In case there is no demand button connected the therapy symbol looks like
this:
.
The demand button is connected to the interface P2 at the rear side of the
pump.
Hint:
It is possible to start a therapy in continuous mode and switch over to
PCA later on (in case the drug is dedicated for use with continuous and PCA
application).
SpacePCA-Chart
If
r
is pressed on the RUN screen, the SpacePCA-Chart is displayed:
The bar represents a time axis, with the points above the axis representing the
number of boli administered and the points below the axis representing the
number of boli refused.
The chart has a 15 minute resolution and shows max of 5 points per 15 min-
utes. Should more then 5 boli be given or refused in this time, the last point will
be turn bold.
Changes to the PCA parameters are displayed as arrowheads at the bottom of
the chart.
3.5 Barcoding
This software version has the barcoding functionality included but initially not
activated. Please contact your local sales representative in case you like to use
barcoding.
The A/D-ratio indicates the
percentage of administered and
demanded boli thus giving an idea
about the effectivity of the therapy.