FlightHawk RF Test Set Calibration
The FlightHawk RF Test Set includes a certificate of calibration at the time of
manufacture, the calibration should be renewed at two year intervals to ensure
accuracy is maintained as indicated in the following specification tables.
It is recommended that an instrument calibration be performed by a certified
service provider at nominal two year intervals. Contact Bird Service Center for
more information. See
FlightHawk-AV Analyzer Specifications
Frequency Range
1 MHz to 6 GHz
Frequency Accuracy
±2.5 ppm
Frequency Resolution Setting
1 kHz
Output Power
–10 dBm
Reflected Amplitude Accuracy
1
–15 dB to 0 dB
–25 dB to –15 dB
–35 dB to –25 dB
±0.4 dB
±1.5 dB
±4.0 dB
Trace Noise Amplitude (IFBW 1kHz)
0.02 dB rms
Single Point Measure Time
1.0 ms
Data Points
51 to 3201
Measure Bandwidth
100 Hz to 30 kHz
Temperature Stability
0.01 dB/°F
0.02 dB/°C
Return Loss Measurement Range
Resolution
0 dB to –60 dB
0.01 dB
VSWR Measurement Range
Resolution
1.0 to 65.0
0.01
Cable Loss Measurement Range
Resolution
0 dB to 30 dB
0.01 dB
DTF Range
0 to 5000 ft.
0 to 1500 m.
Test Port Connector
Impedance
N-type, Female
50 ohms
Connector
USB-C, USB 3.0
Dimensions (L × W × H)
7.7 x 3.6 x 2.4 in.
195 x 90 x 60 mm