FlightHawk RF Test Set
A-4
(Measure Match Mode) using the Max VSWR values from the
table below.
(b) Press the Menu icon. Press RESET to clear all current data and
settings.
(c) Recall previously saved set-up OR select “Measure Match
Mode” to manually input settings.
1) FlightHawk-AV status, calibration result, and tracing data can
be saved to the instrument, and can be recalled to be displayed
on the sweep display area. The parameters for this instrument
setting include the scale, trace, cursor and analysis. These
parameters are saved in the status files. Data in .sta or .cst data
files can be recalled and displayed with the current data trace.
NOTE: .cst files will save the settings and calibration if
the unit was calibrated when the file was saved.
.sta files only save the setting, but not calibration.
a) Push the Menu icon.
b) Push the FILE icon to display the file menu.
c) Push the Save Type to select file type CAL
(cst) or STATUS (sta).
d) Push Load.
e) Select the data file from the list.
f) Push Load.
NOTE: The active and the recalled trace data will be
displayed together in the sweep display area. In the
example, Tr1 is the active trace, the recalled trace data is
indicated with OFF and the file name.
g) Push OFF to remove the recalled trace data.