Chapter 3: Operation
Cutoff Formulas
A cutoff formula calculates a
cutoff value
that is used for classifying samples.
See
Formula Type
on page 70 for instructions on selecting an assay and
accessing the Cutoff formula definition screen.
During data reduction, results are evaluated against the cutoff value (with an
optional greyzone), and each well is assigned a call
POS
(positive),
NEG
(negative), or
EQUIV
(equivocal).
•
One cutoff formula may be defined per assay.
•
If Transformation Formulas are defined, cutoffs are based on the
transformed results. Refer to “Defining Formula” on page 69 for the
order in which formulas are processed.
•
A cutoff formula can consist of a simple numeric value (
1.500
); a well
identifier (
PC
to indicate the criterion for each of the PC replicates, or
PC;x
to indicate the average of the Positive Control replicates); or a
formula combining the two (
NC;x+0.050
).
•
A “greyzone” around the cutoff value can be defined, to indicate
equivocal or indeterminate results.
•
Do not use the
<
or
>
mathematical symbols in a cutoff formula.
Tip:
Choose to print a Column Report to see the greyzone and cutoff
values as well as the equations used to assign calls to samples.
Greyzone Entry
The
greyzone
is a definable area around the cutoff value. Samples that fall
within an area defined by the greyzone (ex. ± 5.0% of the cutoff value)
could be considered equivocal (EQUIV).
E N T E R
G R E Y Z O N E :
0 5
%
Use the
numeric
keys to enter the greyzone percentage.
The valid entry range is from 00 to 99%. An entry of 00% indicates
no greyzone, although a sample equal to the cutoff value will still
receive the EQUIV call.
See
Positive / Negative Calls
on the next page for information on
how calls are assigned.
Bio-Tek Instruments, Inc.
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...