Liquid Testing
|
123
Expected Results:
Since it is somewhat difficult to achieve high pipetting accuracy
when conducting linear dilutions, an R-Square value greater
than or equal to 0.990 is considered adequate.
Repeatability Test
1
Calculate the mean and standard deviation for the five readings
taken above at each concentration. Only one data set needs to be
analyzed for each concentration. The well that shows the most
variation for each concentration is selected for data reduction.
2
For each mean below 2.000 OD, calculate the allowed deviation
using the repeatability specification for a 96-well format of
± 0.5% ± 0.005 OD from 0.000 to 2.000 OD @ 405 nm.
3
The standard deviation for each set of readings should be less
than the allowed deviation.
For example:
Absorbance readings of 1.950, 1.948, 1.955, 1.952, and 1.950 will
result in a mean of 1.951, and a standard deviation of 0.0026. The
mean (1.951) multiplied by 0.5% (1.951 * 0.005) = 0.0098, which,
when added to the 0.005 (0.0098 + 0.005) = 0.0148 OD, which is
the allowable deviation. Since the standard deviation is less than
this value, the reader meets the test criteria.
Repeatability Specification:
For comparison in this test, the following repeatability spec is
applied, using Normal mode, 96-well microplate:
± 0.5% ± 0.005 OD from 0.000 to 2.000 OD @405 nm
ELx800 Operator's Manual
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...