120
|
Chapter 4: Performance Verification and IQ/OQ/PQ Procedures
8.
The standard deviation for each set of readings should be less than
the allowed deviation.
For example: Absorbance readings of 1.950, 1.948, 1.955, 1.952, and
1.950 will result in a mean of 1.951, and a standard deviation of
0.0026. The mean (1.951) multiplied by 0.5% (1.951 * 0.005) = 0.0098,
which, when added to the 0.005 (0.0098 + 0.005) = 0.0148 OD, which
is the allowable deviation. Since the standard deviation is less than
this value, the reader meets the test criteria.
Repeatability Specification:
For comparison in this test, the following repeatability specification
is applied, using the Normal Read Mode and a 96-well microplate.
± 0.5% ± 0.005 OD from 0.000 to 2.000 OD @ 405 nm
Bio-Tek Instruments, Inc.
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...