114
|
Chapter 4: Performance Verification and IQ/OQ/PQ Procedures
To enter the calibration values into KCjunior™:
1.
First define the Absorbance Test Plate to be used for the test. To do
this, open the
Utilities
menu, and select
Diagnostics
|
Universal
Plate Test
.
2.
Click
New Data Sheet
. Click the Help button for guidance.
To enter the calibration values into KC4™:
1.
First define the Absorbance Test Plate to be used for the test. To do
this, open the
System
menu and select
Diagnostics
|
Define
Universal Test Plate|
Add Plate
. Click the Help button for
guidance.
Running the Absorbance Plate Test
IMPORTANT:
Before running the Absorbance Plate
Test, ensure that the reader is not running in
Rapid
Mode. To check this, select
UTIL
READ
and then
cycle through the prompts until
READ IN RAPID
MODE?
is displayed. Choose
NO
for an accurate result.
To run the Absorbance Plate Test from the instrument keypad:
1
Start at the Main Menu and select
UTIL
TESTS
CALPLATE
. The
SELECT TEST
menu will appear:
S E L E C T
T E S T :
S Y S T E M
C H K S U M
C A L P L A T
E
2
Select the appropriate wavelength at the
CALIBRATION FILTER
screen, and press
ENTER
to save the value and continue.
C A L I B R A T I O N
F I L T E R :
4 0 5
4 0 5
4 5 0
4 9 0
* M O R
E
3
When prompted, insert the Test Plate into the plate carrier, and
press the
READ
key to begin the test.
The Calibration Plate Analysis Report (
Figure 19
) will be sent
to a printer when the test is run.
Bio-Tek Instruments, Inc.
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...