Using the Utility Options
|
99
Specifying Data Output and Reporting Options
Plate data can be sent to an attached
printer
or external
computer
.
•
The onboard software provides several different options for report format
and content.
•
Data sent to an external computer has no data reduction applied to it, with
the exception of dual-wavelength subtraction (if defined in the assay). Any
reporting options selected through the onboard software have no effect on
serial output.
1.
To specify data output and reporting options, start at the Main Menu and
select
UTIL
OUTPUT
.
The
REPORT OUTPUT
screen will appear.
R E P O R T
O U T P U T :
B O T H
P R I N T
C O M P U T E R
B O T H
2.
The current output option is displayed on the top line. Select
to send
reports directly to a printer,
COMPUTER
to send data out through the serial
port, or
BOTH
.
Note:
These options have no effect on data output if the instrument
is being controlled by software (such as KCjunior or KC4) running on a
host PC.
3.
Press
ENTER
to continue. The
SELECT PRINTER
screen will appear.
S E
L E C T
P R I N T E R
E P S O
N
E P S O N H P
The ELx800 supports printers using either HP's PCL3 language, such as
the HP DeskJet series, or Epson's LQ language. For the latest list of
compatible printers, consult the Bio-Tek Web site (
),
or call Bio-Tek Instruments' Technical Assistance Center (refer to
Chapter 1
for contact information).
4.
Select
EPSON
or
HP
as appropriate.
ELx800 Operator’s Manual
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...