Specifications
|
7
Ultraviolet/UV Model
The following specifications apply to 96-well, flat- or round-bottom plates.
Wavelength Range:
340 to 750 nm
Filters:
10 nm half-bandwidth interference filters.
User-accessible filter wheel. Up to 5 filters may
be installed on the instrument at one time.
Filters supplied: 340 nm, 405 nm, 450 nm,
490 nm and 630 nm.
•
Optical specifications
for the 400 to 750 nm range (
50-second
read
in
normal read mode):
Absorbance Measurement
Range:
0.000 to 3.000 OD
Accuracy:
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Linearity:
± 1.0% from 0.000 to 2.000 OD @ 405 nm
± 3.0% from 2.000 to 3.000 OD @ 450 nm
Repeatability (STD):
± 0.5% ± 0.005 OD from 0.000 to 2.000 OD @ 405 nm
•
Optical specifications
for the 340 to 400 nm range
(50-second read
in
normal read mode
):
Absorbance Measurement
Range:
0.000 to 3.000 OD
Accuracy:
± 2.0% ± 0.010 OD from 0.000 to 2.000 OD
@ 340 nm
Linearity:
± 2.5% from 0.000 to 2.000 OD @ 340 nm
Repeatability (STD):
± 1.5% ± 0.005 OD from 0.000 to 2.000 OD
@ 340 nm
•
Optical specifications
for the 400 to 750 nm range
(30-second read
in
rapid read mode
):
Absorbance Measurement
Range:
0.000 to 3.000 OD
Accuracy:
± 2.0% ± 0.020 OD from 0.000 to 2.000 OD
@ 405 nm
Linearity:
± 2.0% from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD):
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD
@ 405 nm
ELx800 Operator's Manual
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...