Appendix E: Instructions for Programming a New Assay
Quality Control and Results
Quality Control
For an assay to be valid, all the following criteria must be met:
•
Cutoff as well as Positive and Negative controls must be
included in each run.
•
The OD of the cutoff and the ANA result of the Negative and
Positive Controls should be in the ranges specified on the QC
Certificate.
•
For example: The absorbance of the Positive Control must be
greater than 1.200 OD.
The absorbance of the Negative Control must be less than 0.300
OD.
Note:
If the above criteria are not met, the assay is invalid and
the test should be repeated.
Calculation of the Sample Results
Use the following formula to calculate the ANA result for each sample:
Control or sample OD
x 10 = Control or sample value (U/ml)
Cutoff control OD
9
Expected Values
The normal range was determined on serum from 200 normal adult blood
donors. The cutoff control has been set at a point equivalent to the upper
normal limit, or a cutoff level of 10 U/ml.
The ranges are provided as a guide only. ELISA assays are very sensitive
and capable of detecting small differences in sample populations. It is
recommended that each laboratory determine its own normal range, based
on the population techniques and equipment employed.
ANA Result
Interpretation
<
10.0
Negative
>
10.0
Positive
Bio-Tek Instruments, Inc.
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...