Probe Cleaning
Probe Cleaning
The integrated sample line wash station is located at the bottom of the gauge. The wash station can
introduce either sheath fluid (DI water with additive) or cleaning fluid (from the cleaner bottle) to the
sample line, depending on the process and instrument state.
You can program washes into the sample run during run list setup, and you can specify both the
inside and outside wash times. These washes use filtered (0.2 µm) sheath fluid. The outside of the
sample probe is cleaned by moving the contaminated portion of the sample probe up and down
through the wash station. The inside of the sample line is cleaned by moving the probe up and down
through the wash station while the sample pump is running. This method introduces air and sheath
segments into the sample line and pushes residual sample and debris off the walls and through to
waste. This results in highly effective cleaning and minimal carryover between samples.
The shutdown process uses cleaning fluid to clean the probe and sample line. For more information,
see
Loader Movement
To sample from various types of media, the loader and probe must move in three directions:
n
Left-right (x-axis)
n
In and out (y-axis)
n
Up and down (z-axis)
User Guide | 53
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......