Chapter 7 Creating Experiments and Workspaces
n
After sample boost, the software runs an additional 5 µl of sample to stabilize flow before
acquisition begins.
n
The probe remains in the sample while acquiring the target volume.
n
Return Sample is automatically activated. Excess sample in the line returns to the tube or well
after target volume is acquired.
Important:
Users can set Return Sample to OFF if necessary. However, when using
Volumetric Counting with Return Sample set to OFF, users will experience sample loss. If
sample loss is a concern, run Volumetric Counting with Return Sample set to ON.
Return Sample
When Return Sample is activated, any sample remaining in the sample line after the required number
of events has been acquired is returned to the tube/well.
Flow Rate or Event Rate Settings
The flow rate controls allow you to specify a precise target flow rate for each position, in µl/sec.
The following flow rate options are available:
n
Low
— 0.5 µl/sec (30 µl/min)
n
Normal
— (default) 1.0 µl/sec (60 µl/min)
n
High
— 1.5 µl/sec (90 µl/min)
Note:
The flow rates values are not editable.
Alternatively, you can specify a target event rate, in events/sec (e/s).
During acquisition in standard mode, you can adjust the target flow rate manually (on the instrument)
from 0.1 µl/sec up to 3.5 µl/sec. However, you cannot adjust the flow rate during acquisition in high-
throughput mode.
Note:
In high-throughput mode, all samples must use the same target flow rate.
196 | ZE5 Cell Analyzer and Everest Software
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......