Experiment Builder
Select high-throughput mode to achieve the highest throughput with minimal time between each
sample. After the probe finishes aspirating sample from a position, any programmed wash or agitation
is performed, then the probe immediately moves to the next sample and begins aspirating. Samples
are delivered to the flow cell at a constant rate and sample boost to the flow cell is not utilized.
High-throughput mode cannot be activated for setup (control) wells. Only sample volume limits (not
event limits or gate limits) can be used in high-throughput mode. The Pause After and Return Sample
options are available only in standard sampling mode.
High-throughput
mode
Standard
mode
Number of samples (tubes or wells) in sample line
during run.
Multiple
One
Boost performed before sample acquisition?
No
Yes
Return Sample function allowed?
No
Yes
Pause (between-samples) function allowed?
No
Yes
Wash between samples allowed?
Yes
Yes
Agitation allowed?
Yes
Yes
Event limit allowed?
No
Yes
Gate limit allowed?
No
Yes
Volume limit allowed?
Yes
Yes
Target flow rate can vary per sample?
No
Yes
Target flow rate range
0.5–2.5 µl/sec
(30–150 µl/min)
0.1–3.5 µl/sec
(6–210 µl/min)
Table 21. Differences between high-throughput and standard mode
For more information, see
Selecting Standard or High-Throughput Acquisition on page 199
,
after a Tube or Well on page 200
, and
Returning Sample to a Tube or Well on page 201
.
User Guide | 77
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......