Accessing the Loader
Test Adjustment Event Rate
Tests whether the system can acquire beads at 500 events per
second.
Perform Automatic PMT
Adjustment
Adjusts PMT voltages so that the mean value of the bead population
in all fluorescence channels falls into data channel 128.
Perform Automatic Laser Delay
Adjustment
Automatically sets laser delays so that the relevant signals are
coordinated and attributed to the correct cells.
Collect Validation Data
Adjusts the event rate to 150 events per second and collects a
5,000-event file.
Save Results
Stores data, voltages, and CVs.
Reset Defaults
Resets all voltages so that you can proceed with your experiment.
Table 32. Steps in the QC process, continued
The QC workspace loads automatically regardless of whether it was initiated from the Home window
or Acquisition workspace. When QC completes, the software remains in the Acquisition workspace. To
return to the Home window, click the Home button.
Note:
The Home button on the toolbar is disabled during the QC process. You can return to the
Home window after the QC process completes.
After the QC process is complete, you can access the QC report from the Reports section of the
toolbar.
For more information about reports, see
Quality Control and ZE5-EYE Reports on page 313
Accessing the Loader
To open the loader door and extend the loader
u
Do one of the following:
n
Press the silver sample chamber button on the front of the instrument.
n
Click the Door toggle in the toolbar.
User Guide | 159
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......