Chapter 7 Creating Experiments and Workspaces
5.
In the Plate Metrics area, click inside each text box to change the value within.
The value in each box corresponds to the horizontal or vertical distance between the arrows
nearest the box. For example, the highlighted arrows in the image that follows indicate the
horizontal distance between the centers of two adjacent wells (A1,A2), which is 8.73 mm.
Note:
If the probe does not drop low enough into a well, enter a value for Probe Z Excess
Delta to drive the probe down lower. If the probe hits the well bottom, this information is used
in calibrating the probe Z value.
6.
Edit the dimensions as needed and click Next.
7.
Place the custom device into the sample loader and press Continue.
176 | ZE5 Cell Analyzer and Everest Software
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......