Chapter 2 Hardware Description
The weight of each bottle is continuously monitored so the system can calculate the approximate
remaining run time for each bottle. The system automatically switches between the two sets of sheath
and waste bottles, and displays an alert when less than 1 hr of run time remains.
Sheath Bottles
The two sheath bottles hold DI water for the system. This solution is also used for washing the sample
probe between runs. These bottles have a blue cap. Each holds 4 L of fluid; together, they provide
about 8 hr of continuous run time.
Note:
When refilling these bottles, ensure that the tubing and filter are reinstalled so that they are
in the lower corner of the bottle opposite the cap.
Sheath Additive Bottle
The sheath additive bottle, which has a white cap, contains 250 ml of concentrated balanced salts,
antimicrobial/antifungal agent, and surfactant. This fluid is drip-fed into the internal sheath reservoir on
a continuous basis while the system is running and helps prevent microbial buildup in the system.
28 | ZE5 Cell Analyzer and Everest Software
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......