Configuring Instrument Settings
Item
Function
Editable parameter name
— each parameter name is
editable and is saved with the acquired FCS data. When
changed, this name propagates to plot and histogram axes,
compensation screens, and the Filter Configuration dialog box.
PMT voltage
— allows you to adjust the PMT voltage for the
selected detector, either by entering a value or moving the
slider.
Important:
The voltage adjustments are saved only for
the current experiment. For each new experiment the
values are reset to default, which are set at the most
recent QC baseline. The adjusted PMT voltage values are
included in the FCS file.
Trigger
— use the trigger parameter or parameters to alert the
system to the presence of an event over the threshold. The
trigger plot represents what the electronics are detecting; it
uses log scaling on the x-axis and data in the gray region are
excluded. Raising or lowering the threshold allows you to
exclude unwanted data from acquisition.
Data below the threshold are not saved as part of the FCS file;
only events at or above the threshold are saved.
Select trigger parameter
— the default trigger is forward
scatter (488 nm laser). Everest Software enables data
triggering by up to two unique parameters, either scatter or
fluorescence. The threshold is set using the selected trigger
parameter.
Table 38. PMT and Laser Control Items, continued
User Guide | 259
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......