Error
Possible causes
Troubleshooting steps
Plots created in the workspace do
not match the enabled parameters
Create new plots that match
the enabled parameters (see
), or enable
parameters that match the
plots (see
).
Loader door open
Stop the acquisition process,
close the loader door, and
proceed with running the
experiment.
No events present during QC
process
Probe position not sufficient to
aspirate low fluid levels in QC
bead bottle
Contact Bio-Rad Technical
Support for assistance with
recalibrating the probe
position for the bead station.
Bead bottle empty
Replace bead bottle. See
Bead bottle has been diluted
Replace bead bottle. See
354 | ZE5 Cell Analyzer and Everest Software
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......