Replacing the QC Beads
Replacing the QC Beads
Bio-Rad suggests that you replace the QC beads when the QC calibration bead volume is low (less
than 150 μl), as indicated in the fluidics status. If the fluid level in the calibration bead bottle drops too
low, the system displays air bubble warning messages.
Replacing the QC beads requires the system to re-baseline. Re-baselining establishes a new median
range for CV and PMT voltages in each channel for the new bead lot. Dyed beads have inherent
variation, so calibration must be performed for each lot. The system automatically recalibrates the
instrument after the bead swap is initiated.
Important:
Do not mix the contents of QC bead bottles. When replacing QC beads, do not add
drops from the previous bottle to the new bottle.
To replace the beads
1.
In the System section of the toolbar, click Swap Beads in the Instrument Tools dropdown list.
User Guide | 335
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......