Fluidics Issues
Fluidics Issues
Error
Possible causes
Troubleshooting steps
System suddenly
shuts down
Bulk fluidics bottles empty
Empty the waste and refill DI water bottles as
necessary. See
DI water uptake line and filter
in wrong position
Ensure that the filter on the DI water uptake
line falls into the lower corner of the bottle on
the side away from the cap. Refer to the
figures in
Replacing Sheath Bottles on page
.
Software Issues
Error
Possible causes
Troubleshooting steps
Unable to click the
Play button
Sample back flushing or
previous run not complete
Wait for the process to complete.
Run list error
Validate that everything has been properly
selected when setting up a run list.
Loader door open or sample
plate inserted incorrectly
1. Stop the acquisition process.
2. Open the loader door (if it is not already
open).
3. Ensure that the sample tube rack or plate is
seated correctly.
4. Close the loader door.
5. Proceed with running the experiment.
User Guide | 343
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......