Shutting Down
A message in the Instrument Control Panel indicates that the system is paused.
To resume the system
u
In the System section on the Acquisition workspace toolbar, click Resume System.
Shutting Down
Important:
Do not shut down the ZE5 Cell Analyzer using the main power switch. Refer to
for more information.
Shut down the ZE5 Cell Analyzer system at the end of each day of use. This turns off all lasers; cleans
the sample line, probe, and flow cell with cleaner; depressurizes the fluidics; and logs you out of the
system. The shutdown process is entirely automated and does not require user interaction.
Important:
Always follow the personal protective equipment (PPE) guidelines relevant to your
laboratory’s safety procedures for dealing with ethanol or bleach.
The Shutdown button is located
n
In the Home window
n
In the System Actions section on the Acquisition workspace toolbar
It can also be found in the Login window, if your system administrator has allowed shutdown to occur
when no user is logged in. For more information, see
Specifying Logged Out Settings on page 114
.
User Guide | 161
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......