VDL (E3.1) 06/2020
Page 87/206
8.6
Performance during and after power failures
During a power failure, all controller functions are shut down.
If there is a vacuum and ventilation is required during the duration of the power failure, this is possible via
the emergency ventilation (chap. 9.10.3).
After the power returns, all functions return to the same status the chamber had before power failure.
All settings and set point values remain in the memory during power failure.
•
If the Standby mode had been activated prior to the power failure, the chamber remains off after the
power returns. To operate it you must sign in with a higher authorization level and deactivate the
standby mode.
•
If the Standby mode had been deactivated prior to the power failure, operation continues after the
power returns with previously entered parameters. If the pressure reaches or falls below the required
pressure threshold of 100 mbar, the heater turns on depending on the setpoint.
After the power returns, the controller continues to function in the original
operating mode
it was in pre-
viously before the power failure occurred:
•
Performance after power failure in
Fixed value operation mode
All functions return to the same status the chamber had before power failure. The set-points are im-
mediately resumed.
•
Performance after power failure during
Timer operation
The set-points which had been active at the moment of program start are resumed. Program run con-
tinues.
•
Performance after power failure during
Time program operation
The program is resumed at the point where the interruption occurred with the latest set-points reached
during the program run.
•
Performance after power failure during
Week program operation
The week program continues with the values corresponding to the current time.
Power failure and power return are noted in the
Event list
(chap. 13.5).
If during power failure an
alarm
has occurred (e.g., tolerance range, safety controller), confirm the alarm.
See chap. 16.3.
Summary of Contents for vdl 115
Page 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Page 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Page 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Page 198: ...VDL E3 1 06 2020 Page 198 206 ...
Page 199: ...VDL E3 1 06 2020 Page 199 206 ...
Page 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Page 207: ......