VDL (E3.1) 06/2020
Page 20/206
1.7.4
Safety instructions on operating the vacuum drying oven
Note the following points before starting up the oven:
When loading the chamber and possibly at the moment of unloading, also in the context of
intended use, an explosive mixture may form in the working space. Define a safety area of
at least 1m from the chamber front and ensure active extraction (technical ventilation).
The walkable installation and operating surface of the chamber must be conductive. This
installation and operating surface must be connected to the vacuum drying oven according
to the grounding concept. Cyclic measurements of the equipotential bonding are required.
The operator must ensure an appropriate ventilation of the loading area in front of the ov-
en front prior to commissioning of the chamber.
Ensure that at no time any solvent vapors could enter in the area of the electrical installa-
tion room and the operating panel.
Provide technical ventilation in the area of the vacuum pump stand, particularly in the areas
of the condensate catchpot (when emptying it) and the exhaust air of the vacuum pump.
The personal protective equipment (PPE) of the operating personnel must be ESD pro-
tected.
Only trained personnel with password authorization may work on the VDL vacuum drying
oven.
DANGER
Electrical hazard by water entering the chamber.
Deadly electric shock.
∅
The equipment must NOT become wet during operation, cleaning, or maintenance.
∅
Do NOT install the equipment in damp areas or in puddles.
Set up the equipment in a splash-proof manner.
DANGER
Electrical hazard
due to damage to the equipment
Deadly electric shock
.
∅
Do NOT insert any objects, particularly metallic objects, in louvers or other openings or
slots on the chamber
∅
Do NOT operate the chamber if the housing is damaged.
∅
Do NOT operate the chamber if the power cord is damaged.
Disconnect the chamber from the power supply in case of an obvious malfunction.
CAUTION
Danger of burning when touching the hot inner surfaces during operation.
Burns.
∅
Do NOT touch the inner surfaces or the charging material during and after operation.
Summary of Contents for vdl 115
Page 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Page 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Page 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Page 198: ...VDL E3 1 06 2020 Page 198 206 ...
Page 199: ...VDL E3 1 06 2020 Page 199 206 ...
Page 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Page 207: ......