VDL (E3.1) 06/2020
Page 138/206
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu. The
program starts running.
If instead you press the
Close
icon to exit the menu without taking over the entries, the program will not
start.
Normal display. Information on the bottom of the
screen indicates the currently running program and the
time already passed. The grey bar shows how much
time of the whole time is elapsed. If program duration
has been set to infinite, the grey bar is not displayed.
Heating takes place only when a vacuum according to the pressure threshold of 100 mbar is reached.
Only then start the program or set a tolerance range for the first program section.
18.1.1 Performance during program delay time
During the configured program delay time until program start, the controller equilibrates to the current
setpoints of Fixed value operation mode. Modifications of these setpoints are effective. When the config-
ured moment for program start is reached, the program delay time ends and the program starts running.
18.2 Stopping a running time program
18.2.1 Pausing a running time program
Press the
Program pause
icon to interrupt the program..
The program is paused. The program runtime stops running down, the time display flashes.
There are the following options:
Press the
Program start
icon to continue the program
Press the
Cancelling
icon to cancel the program
18.2.2 Cancelling a running time program
Press the
Program cancelling
icon to cancel the program.
A confirmation prompt is displayed. Press the
Confirm
icon to confirm that the program shall really be
cancelled.
After confirming the message, the controller changes to Fixed value operation mode. Temperature and
humidity will then equilibrate to the setpoints of Fixed value operation mode.
Summary of Contents for vdl 115
Page 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Page 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Page 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Page 198: ...VDL E3 1 06 2020 Page 198 206 ...
Page 199: ...VDL E3 1 06 2020 Page 199 206 ...
Page 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Page 207: ......