VDL (E3.1) 06/2020
Page 54/206
The operator must ensure active extraction before starting up the chamber. Extraction must be provided
during the entire operation of the chamber as well as when handling the condensate catchpot of the
pump. This ensures that solvent vapors never reach unprotected areas or accumulate in an impermissible
manner.
If the technical ventilation fails, the power to the vacuum drying oven must be switched off. It must be
ensured that the device can be switched off by the operator using an explosion-protected device: Pull out
the power plug or operate e.g., a customer’s explosion-protected emergency stop switch.
5.2.3
Air supply (breaking the vacuum) during operation with inert gas
When operating the VDL safety vacuum drying oven with inert gas, correctly follow the technical ventila-
tion measures, as described in the DGUV guidelines 213-850 on safe working in laboratories, issued by
the employers’ liability insurance association (for Germany).
5.3
Equipotential bonding
The accessible installation and operating surface of the chamber must be electrically conductive. The
accessible installation and operating surface of the chamber must be electrically conductive. This installa-
tion and operating surface must be connected to the vacuum drying oven and other equipment (e.g.
Pump module, vacuum pump) according to the grounding concept. Provide cyclic measurements.
For the grounding concept see chap. 6.8.
When inserting or loading objects into the loading area , note that there may be a potential equalization.
The personal protective equipment (PPE) of the operating personnel must be implemented ESD protect-
ed.
5.4
Ambient conditions
•
Permissible ambient temperature during operation: +18 °C /
64 °F
to +32 °C /
90 °F
.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +22 °C +/- 3 °C /
71.6 °F +/-5.4 °F
to which the specified technical data relate.
Deviations from the indicated data are possible for other ambient conditions.
•
Permissible ambient humidity: 70 % r.h. max., non-condensing.
•
Installation height: max. 2000 m /
6562 ft
above sea level.
The maximum permissible ambient temperature of the vacuum pumps delivered by BINDER is 40 °C /
104 °F
.
5.5
Compressed air / inert gas supply for sweeping the area for electrical
equipment and controller housing
Before commissioning the chamber, connect the vacuum drying oven to the compressed air supply (chap.
6.3).
5.6
Fire extinguisher
During operation a fire extinguisher must be available.
Summary of Contents for vdl 115
Page 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Page 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Page 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Page 198: ...VDL E3 1 06 2020 Page 198 206 ...
Page 199: ...VDL E3 1 06 2020 Page 199 206 ...
Page 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Page 207: ......