VDL (E3.1) 06/2020
Page 22/206
If the following precautions are not followed, the vapors resulting from heating of the solvent can ignite on
the hot walls of the inner chamber
Required measures for operation with solvent-containing materials that can form an
explosive mixture with air:
•
Sweep the area for electrical equipment and controller housing with compressed air or inert
gas with maximum overpressure for at least 10 minutes. Subsequent sweeping of the area
for electrical equipment and controller housing with an overpressure of at least 25 Pa (rec-
ommendation: 40 Pa) must take place during the entire operating cycle and must be moni-
tored at the manometer on the chamber front.
•
Only after completing this sweeping time, put the chamber in operation.
•
Take the auto-ignition temperature from the safety specifications of the solvent. In the case
of solvent mixtures, use the auto-ignition temperature of the material with the lowest auto-
ignition temperature. Make sure that the auto-ignition temperature is above 200 °C.
•
Enter the temperature set-point (drying temperature) on the controller.
•
Set the safety controller (temperature limiter class 2). Recommended setting: safety con-
troller mode „Limit“, safety controller value approx. 5 °C above the set-point.
•
During the drying process, use only those solvents for which the auto-ignition temperature
has been entered.
•
Make sure that the technical ventilation is active. Introduce the drying material.
•
Start the drying process. The heating unit is only released when a pressure threshold of
100 mbar is reached.
•
Make sure that sweeping the area for electrical equipment and controller housing with at
least 25 Pa overpressure (see manometer display) takes place during the entire operating
cycle. Recommended value: 40 Pa.
•
The duration of the drying process can be determined by means of the pressure display on
the controller. When the pressure drops to the set-point, the drying process is finished.
•
Make sure that the technical ventilation is active. After termination of the drying process,
ventilate the vacuum drying oven with ambient air or inert gas. Unload the loading material
and turn off the oven.
•
Sweep the area for electrical equipment at least 10 minutes with compressed air or inert
gas (recommended).
•
Make sure that the technical ventilation is active when emptying the condensate catchpot
of the vacuum pump.
•
Make sure that sweeping the area for electrical equipment and controller housing with at
least 25 Pa overpressure (see manometer display) is active when emptying the conden-
sate catchpot of the vacuum pump. Recommended value: 40 Pa.
•
Before starting a new drying process with a different solvent, check again the auto-ignition
temperature at the safety specifications of the solvent. In the case of solvent mixtures, use
the auto-ignition temperature of the material with the lowest auto-ignition temperature.
Make sure that the auto-ignition temperature is above 200 °C.
Summary of Contents for vdl 115
Page 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Page 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Page 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Page 198: ...VDL E3 1 06 2020 Page 198 206 ...
Page 199: ...VDL E3 1 06 2020 Page 199 206 ...
Page 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Page 207: ......