Fault description
Possible cause
Required measures
Controller
(continued)
Program duration longer than
programmed.
Tolerances have been pro-
grammed.
For rapid transition phases, do NOT
program tolerance limits in order to
permit maximum heating, evacuat-
ing, or ventilation speed.
Program keeps the last program
setpoint constant while in setting
“ramp”.
Program line with setting
“ramp” is incomplete.
When programming with setting
“ramp”, define the end value of the
desired cycle by adding an addition-
al section with a section time of at
least one second.
Ramp temperature transitions are
only realized as steps.
Setting “step” has been se-
lected.
Select setting “ramp”.
26.3
Maintenance, Service
26.3.1 Safety instructions on maintenance work
DANGER
Fire and explosion hazard due to explosive atmosphere during maintenance .
Serious injury or death from burns and / or explosion pressure.
Before performing maintenance work, make sure that there is no explosive atmosphere
in the installation area of the chamber or inside the chamber.
∅
Do NOT perform any maintenance or tests on the chamber in potentially explosive are-
as s.
Conduct the explosion protection test when performing the annual maintenance.
∅
Do NOT put the chamber into operation if it did not pass the explosion protection test.
Before commissioning, ensure that the system is correctly and completely grounded.
DANGER
Electrical hazard during live maintenance work.
Deadly electric shock.
Disconnect the chamber before conducting maintenance work. Pull the power plug.
∅
Do NOT remove the rear panel of the chamber.
Make sure that general maintenance work will be conducted by licensed electricians
with additional skills in explosion protection (ATEX) or experts authorized by BINDER.
Summary of Contents for vdl 115
Page 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Page 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Page 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Page 198: ...VDL E3 1 06 2020 Page 198 206 ...
Page 199: ...VDL E3 1 06 2020 Page 199 206 ...
Page 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Page 207: ......