KBF P + KBF LQC + KBWF (E6) 11/2016
page 96/189
Submenu “Various”.
Select the field “Humidity range”.
“Humidity range” entry menu.
Enter the desired value for the humidity range and
press the Confirm icon.
Setting range: 5% r.H. to 20% r.H.
There are the following possibilities:
Press the
Confirm
icon to take over the entries and exit the menu.
Press the
Close
icon to exit the menu without taking over the entries.
If there are actual values outside the tolerance range the following information icons for the corresponding
parameter are displayed:
Icon
Signification
Information
Temperature range
The temperature value is outside the tolerance range
Humidity range
The humidity value is outside the tolerance range
If the condition persists, an alarm is triggered after the configured interval (“range alarm delay”). It is visu-
ally indicated in Normal display. If the alarm buzzer is activated (chap. 12.5) there is an audible alert. The
zero-voltage relay alarm output (option, chap. 21.5) switches to transmit the alarm. The alarm is shown in
the list of active alarms (chap. 12.3).
12.5 Activating / deactivating the audible alarm (alarm buzzer)
Path:
Main menu > Settings > Chamber
“Chamber” submenu.
In the field “Audible alarm” select the desired setting “off“
or “on” and press the
Confirm
icon.
Summary of Contents for KBF P 240-UL
Page 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Page 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Page 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Page 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Page 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Page 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Page 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Page 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Page 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...