Radiation in a range of +/- 100° around the 90° axis of incidence is weighted almost equally with a factor
between 1.0 and 0.9. Only with greater angles, weighting of the radiation decreases, technically caused by
the sensor foot.
Comparison of different sensor types
Planar sensor
Spherical sensor
(sensor foot at 270°)
Ideal spherical sensitivity characteristics
Figure 32: Comparison of the spatial sensitivity characteristics
of planar and spherical sensors
If a sample is irradiated with light at an angle of incidence of 45°, the
spherical sensor assumes the impinging light quantity as large as the
amount of light that would impinge with vertical irradiation (factor 1).
Since this is true for a sample with spatial extension, the error is zero
in this case.
A planar sensor instead would take into account the cosine of the an-
gle of incidence: cos 45 = 0.71. But due to the spatial extension of the
sample, no correction is necessary. The planar sensor therefore calcu-
lates the amount of light which impinges on the sample by a factor of
0.71 less than the light quantity that would impinge with vertical irradia-
tion. Thus, the light really impinging on the sample surface will be cal-
culated only at 71% of the real value.
Deviation from the vertical: 0°
Light
source
In case of exclusive use of entirely flat samples without spatial extension (e.g. spreaded sam-
ples, film), an overrating of the light which really impinges on the plane surface is possible.
Please contact BINDER INDIVIDUAL customized solutions.
If an independent reference measuring device shall be used to directly compare the light inten-
sities, it must bear the same spatial sensitivity characteristics (spherical) as the sensors of the
BINDER system.
45°
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
90
100
110
120
130
140
150
160
170
180
190
200
210
220
230
240
250
260
270
280
290
300
310
320
330
340
350
0
10
20
30
40
50
60
70
80
Summary of Contents for KBF P 240-UL
Page 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Page 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Page 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Page 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Page 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Page 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Page 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Page 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Page 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...