KBF P + KBF LQC + KBWF (E6) 11/2016
page 147/189
21. Options
21.1 Communication software APT-COM™ 3 DataControlSystem (option)
The chamber is regularly equipped with an Ethernet interface (4) that can connect the BINDER communi-
cation software APT-COM™ 3 DataControlSystem. The actual temperature and humidity values are given
at adjustable intervals. Programming can be performed graphically via PC. Up to 30 chambers with RS
422 interface can be cross-linked. The MAC Address is indicated in the “Device info” controller menu
(chap. 15.5.2.2). For further information on networking please refer to the operating manual of the
BINDER communication software APT-COM™ 3.
KBF-LQC:
The integrated values of VIS and UVA light are documented GLP/GMP compliant by the APT-
COM™ 3 software. The documentation of these light values takes place under the same preconditions
regarding 21CFR11 as the documentation of temperature and humidity. The user thus receives an overall
solution perfectly adapted to his chamber.
21.2 RS485 interface (option)
With this option, the chamber is equipped with an additional 2-wire RS485 serial interface (7) that can
connect the BINDER communication software APT-COM™ 3 DataControlSystem. The actual temperature
and humidity values are given at adjustable intervals. For further information, please refer to the operating
manual of the BINDER communication software APT-COM™ 3.
21.3 Data logger kits (option)
BINDER Data Logger Kits offer an independent long-term measuring system for temperature and humidi-
ty, available for different temperature ranges. According to the selected kit, the Data Logger can measure
and record also the ambient temperature and humidity values via a second multi-function sensor.
BINDER Data Loggers are equipped with a keyboard and a large LCD display, alarm functions and a real-
time function. Measurement data are recorded in the Data Logger and can be read out after the meas-
urement via the RS232 interface of the Data Logger. It offers a programmable measuring interval and
permits storing up to 64000 measuring values. Reading out is done with the Data Logger evaluation soft-
ware. You can give out a combined alarm and status protocol directly to a serial printer.
Data Logger Kit TH 70:
Multi-function sensor for chamber temperature and humidity: Temperature range
-40 °C /
-40 °F
up to 70 °C /
158 °F
, humidity range 0% r.H. up to 100% r.H.
Data Logger Kit TH 70/70:
Multi-function sensor for chamber temperature and humidity: Temperature
range -40 °C /
-40 °F
up to 70 °C /
158 °F,
humidity range 0% r.H. up to 100% r.H. Multi-function sensor
for ambient temperature and humidity: Temperature range -40 °C /
-40 °F
up to 70 °C /
158 °F,
humidity
range 0% r.H. up to 100% r.H.
For detailed information on installation and operation of the BINDER Data Logger, please
refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the
manufacturer, supplied with the data logger.
Summary of Contents for KBF P 240-UL
Page 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Page 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Page 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Page 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Page 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Page 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Page 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Page 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Page 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...