KBF P + KBF LQC + KBWF (E6) 11/2016
page 169/189
Chamber size
240
720
Climatic data
(with humidity) for KBWF
Temperature range without light cassettes
°C /
°F
+10 to +70 /
50 to 158
+10 to +70 /
50 to 158
Temperature range with light cassettes, with illumination
°C /
°F
+10 to +60 /
50 to 140
+20 to +60 /
68 to 140
Temperature fluctuation with illumination
± K
0.1 to 1.0
0.1 to 0.5
Temperature uniformity (variation) with illumination
± K
0.5 to 1.0
1.0 to 1.5
Humidity range without light cassettes
% r.H.
10 to 80
10 to 80
Humidity range with light cassettes, with illumination
% r.H.
10 to 75
10 to 75
Humidity fluctuation with illumination
± % r.H.
≤ 2
≤ 2,5
Illumination data per light cassette
ICH compliant illumination for photo stability testing
Lux
7500
9000
UVA W/m
2
1.1
1.5
Daylight fluorescent tubes
Lux
9000
13000
W/m
2
24
38
Fluora
®
growth lamps
Lux
7500
10500
W/m
2
23
36
Arabidopsis lamps
Lux
11000
14000
W/m
2
32
43
Electrical data
(model versions KBFP-230V, KBFLQC-230V, KBWF-230V)
IP system of protection acc. to EN 60529
20
20
Nominal voltage (+/-10%) at 50 Hz power frequency
V
200-230
200-230
Current type
1N~
1N~
Power plug
shock proof plug
Nominal power
kW
2.40
3.50
Installation category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Over-current release category B, 2 poles
Amp
16
16
Different electrical data for UL chambers constructed for the USA and Canada
(model versions KBFP240UL-240V, KBFP720UL-240V, KBFLQC240UL-240V, KBFLQC720UL-240V)
Nominal voltage (+/-10%)
at 50 Hz power frequency
V
200-240
200-240
at 60 Hz power frequency
V
200-240
200-240
Current type
2~
2~
Power plug
NEMA
6-20P
6-20P
Environment-specific data
Noise level (mean value)
dB (A)
53
59
Energy consumption at 40 °C /
104 °F
and 75 % r.H.
Wh/h
760
1850
Filling weight of refrigerant R 134A (GWP 1300)
kg
0.575
0.800
Illumination data:
Average value, measured at +22 °C +/- 3 °C /
71.6 °F +/- 5.4 °F
with a spherical sen-
sor (+/-10%) by 12 cm /
4.7 in
below the light cassette. The values given in W/m
2
refer to global radiation.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3°C /
71.6 °F +/- 5.4 °F
and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER factory standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifica-
tion and may affect the humidity accuracy.
Summary of Contents for KBF P 240-UL
Page 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Page 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Page 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Page 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Page 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Page 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Page 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Page 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Page 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...